Home

Products:
3-4 Crucible Evaporator

9 Crucible Evaporator

Metal Evaporator

MBE-LEED

Low Temperature Head

Ion Source

L. Proj. Pattern

Organic Molecules


Services:

Vacuum systems

Plasma etching

Consulting


Price List & References

Arrival

Low Temperature Measurement Head

Low Temperature Measurement Head

Together with a He flow cryostat the measurement head can carry out four- pole measurements below 10K with electrodes applying a very small force to the layer.

Features:

  • Temperature of the sample 6.6K achieved
  • Freely available electrode material (Au coated tungsten, Pt, Au..)
  • Depending on the size of the electrodes forces of less than 1x10-6N to 0.1N can be applied
  • Omicron-compatible sample holder
  • Low-temperature piezo actuator for very smooth approach (like STM techniques)
  • Final temperature (10K) after 1.2 hours of cooling down

Low Temperature Head

Fig. 1: Measurement head. In sample position a Si temperature sensor is mounted to measure the true temperature

Temperature curve

Fig. 2: Temperature behaviour of the cryostat and sample. Consumption of LHe 2.2l/hour.

The measurement head is completely shielded by the cap of the cryostat.

The (semiconductor) sample can be prepared at different preparation stage by direct heating to 1200°C.








kentax UHV equipment, www.kentax.de, info@kentax.de,