Home
Products:
3-4 Crucible Evaporator
9 Crucible Evaporator
Metal Evaporator
MBE-LEED
Low Temperature Head
Ion Source
L. Proj. Pattern
Organic Molecules
Services:
Vacuum systems
Plasma etching
Consulting
Price List & References
Arrival
|
 |
Low Temperature Measurement Head |
Together with a He flow cryostat the measurement head can carry out four- pole measurements below 10K with electrodes applying a very small force to the layer.
Features:
- Temperature of the sample 6.6K achieved
- Freely available electrode material (Au coated tungsten, Pt, Au..)
- Depending on the size of the electrodes forces of less than 1x10-6N to 0.1N can be applied
- Omicron-compatible sample holder
- Low-temperature piezo actuator for very smooth approach (like STM techniques)
- Final temperature (10K) after 1.2 hours of cooling down
 |
Fig. 1: Measurement head. In sample position a Si temperature sensor is mounted to measure the true temperature |
 |
Fig. 2: Temperature behaviour of the cryostat and sample. Consumption of LHe 2.2l/hour. |
The measurement head is completely shielded by the cap of the cryostat.
The (semiconductor) sample can be prepared at different preparation stage by direct heating to 1200°C.
|